Source: Fresenius Umwelttechnik GmbH
The BioBasic Bionix online gas analyzer monitors biomethane (RNG) quality at the grid injection point and reports CH₄, CO₂, O₂ and H₂S once per second, while a process gas chromatograph delivers a value only every few minutes. CH₄ and CO₂ are measured by dual-beam NDIR, H₂S by NDUV and O₂ electrochemically. In a six-day field comparison, the mean deviation from the process GC was below 0.3 vol.% for CH₄, and the analyzer showed changes about eleven minutes earlier on average, depending on GC cycle and sample line. It complements the GC at a significantly lower cost. Automatic, error-monitored calibration, a color touchscreen, USB data logging, a Bluetooth app and 4–20 mA, RS232 and fieldbus outputs are standard or optional. — , Herten, Germany